Online Oxygen Purity Analyzer
Price
Request a Quote
Stock
In Stock
The OMD-775 is designed to measure elevated oxygen levels (21% – 100%) for online applications.
The unit combines an advanced set of electronic features with our industry leading precision electrochemical oxygen sensor technology.
The analyzer also has built in temperature and barometric pressure compensation, the result is a highly reliable and cost effective design with an easy-to-use user interface.
The analyzer comes standard with a 0 – 100% full scale range with a resolution of 0.01%.
The analyzer can also be configured by the end user in the field for suppressed ranges of 50 – 100%, 80 – 100% and 90 – 100%.
The analyzer comes with the ability to auto-range through all four ranges, this is accomplished with the choice of a voltage or current analog output.
The analyzer offers the user 2 different digital communication options, both of which are bi-directional.
These come in the form of MODBUS RS485 ASCII or RS232.
Alarm functionality is performed by 2 fully adjustable form C non-latching relay contacts.
These can be configured as NO or NC and can be set as HIGH or LOW with optional delay mode.
A power failure alarm is also standard and comes as a form C non-latching relay contact.
The oxygen sensor used in the OMD-775 is based on the galvanic electrochemical fuel cell principal and is compensated for both pressure and temperature giving the user added accuracy.
All oxygen sensors are manufactured in house by Southland Sensing Ltd. under a strict quality program.
The analyzer is a cost-effective, yet similarly performing, alternative to paramagnetic oxygen sensors packaged in a 19” rack mount for easy installation and operation.
Features- 0/50/80/90 - 100% Suppressed Ranges
- Advanced Digital Temperature Compensation
- Integral Barometric Pressure Compensation
- Bi-Directional Communication Link
- Extensive Electronic Features & Outputs
- 19” Rack Mount Configuration
- Electrochemical Sensor Technology
Specifications
Brand
Accuracy
- < ± 1 % of Full Scale Range
Alarms
- Adjustable Relay Contacts
- Power Fail Relay Contact
Analysis Range
- 0/50/80/90 - 100 %
Calibration
- Periodically
Communication
- Bi-Directional RS-232
- Modbus RS-485 ASCII
Dimensions
- 19.0 x 5.125 x 7.75 inch
Display
- Large with Backlight
Enclosure
- Brushed Stainless Steel
Flow Sensitivity
- 0.5 - 5 SCFH
Gas Connections
- 1/8” Compression Tube
Output (Analog)
- 0-1 Vdc, 0-10 Vdc
- Isolated 4-20 mA
Power
- 100-240 Vac
Pressure
- Inlet, 0 - 50 psig
Range ID
- 4-20 mA analog output
- 0-1 Vdc analog output
Response Time
- T90 in 30 seconds
Sample System
- Flow Control, 4-Way sample / Flow Indicator
Sensor
- PO₂-1120 Percent O₂ Sensor
Sensor Life
- 20 - 25 months @ 100 % O₂
Temperature
- 0° to 50° C
Temperature Compensation
- Digital
Warranty
- 12 months Analyzer & Sensor
Weight
- 13.0 lbs
- Air Liquefaction and Separation
- Gas Purity Certification
- Oxygen Monitoring of Medical Gases
- Ozone Generation PSA, VPSA
- Measurement in Wastewater Treatment Plants