Portable Trace Hydrogen Sulfide Analyzer H2S-780

Brand: Southland Sensing Ltd

The H2S-780 Portable Hydrogen Sulfide Analyzer combines a rugged portable design, integral sample system and updated electronics with SSO2’s precision H2S sensors. 

The result is a highly reliable and cost-effective design with easy to use interface.

The analyzer can come with three sets of ranges: The H2S-1x Sensor has ranges 0 – 10 ppm, 0 – 50 ppm, 0 – 100 ppm, 0 – 200 ppm.  

The H2S-2x Sensor has ranges: 0 – 100 ppm, 0 – 500 ppm, 0 – 1000 ppm, 0 – 2000 ppm. The H2S-3x has 0 – 1000 ppm, 0 – 2000 ppm, 0 – 5000 ppm, and 0 – 10000 ppm.

The sensor is based on the galvanic electrochemical fuel cell principal and is manufactured in house under a strict quality procedure.  

The sensors are self contained and minimal maintenance is required – no need to clean electrodes or add electrolyte. 

 The precision sensors offer excellent performance, accuracy and stability while maximizing the expected life.

The analyzer comes with a built in sample system that includes a flow control knob and flow indicator.  

Gas inlet is made via 1/8″ Compression tube fitting.

The electronics are feature packed with an easy-to-use menu interface.  

The highlight of the advanced set of electronics includes a USB output for data logging. 

 Data logging can be done via a removable USB flash drive and data can be taken in 1 minute intervals.  

An 8 GB flash drive can store data for 50+ years.
Features
  • Designed for the Natural Gas Industry
  • Large Backlit Display w/ User Friendly Menu
  • Data Log via Removable USB Flash Drive
  • Integral Flow Control Valve & Flow Indicator
  • Continuous Analysis during Charging
  • Electrochemical Sensor Technology

Specifications

Accuracy

  • < ± 1 % of Full Scale Range

Analysis Range

  • Multiple up to 10000 ppm

Battery Indicator

  • Integrated into Large Display

Calibration

  • Periodically

Data Logging

  • Removable USB Flash Drive

Dimensions

  • 10.9 x 10.0 x 4.9 inch

Display

  • Large with Backlight

Enclosure

  • Brushed Stainless Steel

Flow Sensitivity

  • 0.25 - 5 SCFH

Gas Connections

  • 1/8” Compression Tube

Output (Analog)

  • 0-1 Vdc

Power

  • Rechargeable Battery 100-240 Vac AC Adapter

Pressure

  • Inlet, 0 - 50 psig

Response Time

  • T90 in 45 seconds

Sample System

  • Flow Control, Flow Indicator

Sensor

  • Precision Electrochemical

Sensor Life

  • 12 - 36 months

Temperature

  • 0° to 50° C

Temperature Compensation

  • Integral

Warranty

  • 12 months

Weight

  • 11.25 lbs

  • Natural Gas Extraction & Pipelines 
  • Natural Gas Processing 
  • Landfill Gas Monitoring 
  • Biogas before And After The H2S Scrubber 
  • And Many Other Industrial Applications

Documentation